Accéder directement au contenu Accéder directement à la navigation
Communication dans un congrès

Impact of bipolar electrodes contact on fractionation index measurement.

Abstract : : Signals such as Complex Fractionated Atrial Electrograms (CFAE) are tracked during ablation procedures to locate the arrhythmical substrate regions. Most of CFAE classification tools use fractionation indexes. However, recordings from intracardiac catheter depend on electrode contact quality. This paper investigates the impact of electrode contact area on fractionation indexes. It is assessed through three kinds of arrhythmical activations resulting from a numerical simulation of a small piece of the cardiac tissue. Bipolar electrograms are extracted corresponding to 25 different contact areas and fractionation indexes (Shannon entropy, non linear energy operator and maximum peak ratio) are computed. Results yield that the Shannon entropy offers a good potential discrimination between arrhythmic scenarios and is less sensitive to the electrode contact variation.
Liste complète des métadonnées

https://hal-univ-bourgogne.archives-ouvertes.fr/hal-00872243
Contributeur : Sabir Jacquir <>
Soumis le : vendredi 11 octobre 2013 - 14:53:38
Dernière modification le : vendredi 17 juillet 2020 - 14:54:05

Identifiants

Citation

Nicolas Navoret, Sabir Jacquir, Gabriel Laurent, Stephane Binczak. Impact of bipolar electrodes contact on fractionation index measurement.. Conf Proc IEEE Eng Med Biol Soc, Jun 2013, Osaka, Japan. pp.4211-4214, ⟨10.1109/EMBC.2013.6610474⟩. ⟨hal-00872243⟩

Partager

Métriques

Consultations de la notice

558